ASTM F2477-19

ASTM F2477-19

May 2019
Standard Cancelled

Standard Test Methods for <emph type="bdit">in vitro</emph> Pulsatile Durability Testing of Vascular Stents

This test method covers the procedure for determining the durability of ballon-expandable and self- expanding metal or alloy vascular stents. Tests are performed by exposing specimens to physiologically relevant diametric distention levels using hydrodynamic pulsatile loading. Specimens should have been deployed into a mock or elastically simulated vessel prior to testing. The test methods are valid for determining stent failure due to typical cyclic blood vessel diametric distention and include physiological pressure tests and diameter control tests. These do not address other modes of failure such as dynamic bending, torsion, extension, crushing, or abrasion. Test apparatus include a pressure measurement system, dimensional measurement devices, a cycle counting system, and a temperature control system.

Additional informations

1.1 These test methods cover the determination of the durability of a vascular stent by exposing it to physiologically relevant diametric distension levels by means of hydrodynamic pulsatile loading. This testing occurs on a stent test specimen that has been deployed into a mock (elastically simulated) vessel. The typical duration of this test is 10 years of equivalent use (at 72 beats per minute), or at least 380 million cycles.

1.2 These test methods are applicable to balloon-expandable and self-expanding stents fabricated from metals and metal alloys. It does not specifically address any attributes unique to coated stents, polymeric stents, or biodegradable stents, although the application of this test method to those products is not precluded.

1.3 These test methods do not include recommendations for endovascular grafts (“stent-grafts”) or other conduit products commonly used to treat aneurismal disease or peripheral vessel trauma or to provide vascular access, although some information included herein may be applicable to those devices.

1.4 These test methods are valid for determining stent failure due to typical cyclic blood vessel diametric distension. These test methods do not address other modes of failure such as dynamic bending, torsion, extension, crushing, or abrasion.

1.5 These test methods do not address test conditions for curved mock vessels.

1.6 These test methods do not address test conditions for overlapping stents.

1.7 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

1.8 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.

1.9 General Caveat—This document contains guidance for testing as is currently carried out in most laboratories. Other testing techniques may prove to be more effective and are encouraged. Whichever technique is used, it is incumbent upon the tester to justify the use of the particular technique, instrument, and protocol. This includes the choice of and proper calibration of all measuring devices. Deviations from any of the suggestions in this document may be appropriate but may require the same level of comprehensive justification that the techniques described herein will require.

1.10 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.


Main informations

Collections

ASTM american standards

Publication date

May 2019

Number of pages

0 p.

Reference

ASTM F2477-19

Classification index

F2477-19
Sumary
Standard Test Methods for in vitro Pulsatile Durability Testing of Vascular Stents

This test method covers the procedure for determining the durability of ballon-expandable and self- expanding metal or alloy vascular stents. Tests are performed by exposing specimens to physiologically relevant diametric distention levels using hydrodynamic pulsatile loading. Specimens should have been deployed into a mock or elastically simulated vessel prior to testing. The test methods are valid for determining stent failure due to typical cyclic blood vessel diametric distention and include physiological pressure tests and diameter control tests. These do not address other modes of failure such as dynamic bending, torsion, extension, crushing, or abrasion. Test apparatus include a pressure measurement system, dimensional measurement devices, a cycle counting system, and a temperature control system.

Replaced standards (1)
ASTM F2477-07(2013)
March 2013
Standard Cancelled
Standard Test Methods for <emph type="bdit">in vitro</emph> Pulsatile Durability Testing of Vascular Stents

This test method covers the procedure for determining the durability of ballon-expandable and self- expanding metal or alloy vascular stents. Tests are performed by exposing specimens to physiologically relevant diametric distention levels using hydrodynamic pulsatile loading. Specimens should have been deployed into a mock or elastically simulated vessel prior to testing. The test methods are valid for determining stent failure due to typical cyclic blood vessel diametric distention and include physiological pressure tests and diameter control tests. These do not address other modes of failure such as dynamic bending, torsion, extension, crushing, or abrasion. Test apparatus include a pressure measurement system, dimensional measurement devices, a cycle counting system, and a temperature control system.

Standard replaced by (1)
ASTM F2477-23
February 2023
Standard Cancelled
Standard Test Methods for <emph type="ital"> in vitro</emph> Pulsatile Durability Testing of Vascular Stents and Endovascular Prostheses

This test method covers the procedure for determining the durability of ballon-expandable and self- expanding metal or alloy vascular stents. Tests are performed by exposing specimens to physiologically relevant diametric distention levels using hydrodynamic pulsatile loading. Specimens should have been deployed into a mock or elastically simulated vessel prior to testing. The test methods are valid for determining stent failure due to typical cyclic blood vessel diametric distention and include physiological pressure tests and diameter control tests. These do not address other modes of failure such as dynamic bending, torsion, extension, crushing, or abrasion. Test apparatus include a pressure measurement system, dimensional measurement devices, a cycle counting system, and a temperature control system.

What is the Redline format?
The Redline + service - standards comparator allows you to easily and simply identify major changes between the current standard and its last canceled version.

At a glance, you will be able to identify the additions, deletions or modifications to a text, table, figure and formula.
At a glance, you will be able to identify the additions, deletions or modifications to a text, table, figure and formula

The Redlines + service is offered to you on the collection of French standards in force, in French language and in HTML and PDF format.

For an overview of the service, click on View a standard in redline format
Need to identify, monitor and decipher standards?

COBAZ is the simple and effective solution to meet the normative needs related to your activity, in France and abroad.

Available by subscription, CObaz is THE modular solution to compose according to your needs today and tomorrow. Quickly discover CObaz!

Request your free, no-obligation live demo

I discover COBAZ