ISO 80601-2-84:2023

ISO 80601-2-84:2023

November 2023
International standard Current

Medical electrical equipment - Part 2-84: Particular requirements for the basic safety and essential performance of ventilators for the emergency medical services environment

This document applies to the basic safety and essential performance of an EMS ventilator in combination with its accessories, hereafter also referred to as ME equipment: ¾     intended for patients who need differing levels of support from artificial ventilation including ventilator-dependent patients; ¾     intended to be operated by a healthcare professional operator; ¾     intended for use in the EMS environment; and ¾     intended for invasive or non-invasive ventilation. NOTE 2     An EMS ventilator can also be used for transport within a professional healthcare facility. An EMS ventilator is not considered to use a physiologic closed loop-control system unless it uses a physiological patient variable to adjust the  artificial ventilation therapy settings. This document is also applicable to those accessories intended by their manufacturer to be connected to the ventilator breathing system, or to an EMS ventilator, where the characteristics of those accessories can affect the basic safety or essential performance of the EMS ventilator. NOTE 3     If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant. Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in IEC 60601‑1:2005+AMD1:2012+AMD2:2020, 7.2.13 and 8.4.1. NOTE 4     Additional information can be found in IEC 60601-1:2005+AMD1:2012+AMD2:2020, 4.2. This document does not specify the requirements for the following: NOTE 5     See ISO/TR 21954 for guidance on the selection of the appropriate ventilator for a given patient.  ¾     ventilators or accessories intended for ventilator-dependent patients in critical care applications, which are given in ISO 80601-2-12. ¾     ventilators or accessories intended for ventilator-dependent patients in the home healthcare environment, which are given in ISO 80601-2-72. ¾     ventilators or accessories intended for anaesthetic applications, which are given in ISO 80601‑2‑13. ¾     ventilators or accessories intended for ventilatory support equipment (intended only to augment the ventilation of spontaneously breathing patients), which are given in ISO 80601‑2-79 and ISO 80601-2-80. ¾     obstructive sleep apnoea therapy ME equipment, which are given in ISO 80601‐2‐70. ¾     user-powered resuscitators, which are given in ISO 10651‐4. ¾     gas-powered emergency resuscitators, which are given in ISO 10651‐5. ¾     continuous positive airway pressure (CPAP) ME equipment. ¾     high‐frequency jet ventilators (HFJVs), which are given in ISO 80601-2-87. ¾     high‐frequency oscillatory ventilators (HFOVs) REF _Ref69393406 \r \h  \* MERGEFORMAT [44] 08D0C9EA79F9BACE118C8200AA004BA90B02000000080000000D0000005F00520065006600360039003300390033003400300036000000 , which are given in ISO 80601-2-87. NOTE 6      An EMS ventilator can incorporate high-frequency jet or high-frequency oscillatory ventilation-modes. ¾     respiratory high-flow therapy equipment, which are given in ISO 80601-2-90. NOTE 7      An EMS ventilator can incorporate high-flow therapy operational mode, but such a mode is only for spontaneously breathing patients. ¾     oxygen therapy constant flow ME equipment. ¾     cuirass or “iron‐lung” ventilators.

Main informations

Collections

International ISO standards

Publication date

November 2023

Number of pages

138 p.

Reference

ISO 80601-2-84:2023

ICS Codes

11.040.10   Anaesthetic, respiratory and reanimation equipment

Print number

1
Sumary
Medical electrical equipment - Part 2-84: Particular requirements for the basic safety and essential performance of ventilators for the emergency medical services environment

This document applies to the basic safety and essential performance of an EMS ventilator in combination with its accessories, hereafter also referred to as ME equipment:

¾     intended for patients who need differing levels of support from artificial ventilation including ventilator-dependent patients;

¾     intended to be operated by a healthcare professional operator;

¾     intended for use in the EMS environment; and

¾     intended for invasive or non-invasive ventilation.

NOTE 2     An EMS ventilator can also be used for transport within a professional healthcare facility.

An EMS ventilator is not considered to use a physiologic closed loop-control system unless it uses a physiological patient variable to adjust the  artificial ventilation therapy settings.

This document is also applicable to those accessories intended by their manufacturer to be connected to the ventilator breathing system, or to an EMS ventilator, where the characteristics of those accessories can affect the basic safety or essential performance of the EMS ventilator.

NOTE 3     If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant.

Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in IEC 60601‑1:2005+AMD1:2012+AMD2:2020, 7.2.13 and 8.4.1.

NOTE 4     Additional information can be found in IEC 60601-1:2005+AMD1:2012+AMD2:2020, 4.2.

This document does not specify the requirements for the following:

NOTE 5     See ISO/TR 21954 for guidance on the selection of the appropriate ventilator for a given patient. 

¾     ventilators or accessories intended for ventilator-dependent patients in critical care applications, which are given in ISO 80601-2-12.

¾     ventilators or accessories intended for ventilator-dependent patients in the home healthcare environment, which are given in ISO 80601-2-72.

¾     ventilators or accessories intended for anaesthetic applications, which are given in ISO 80601‑2‑13.

¾     ventilators or accessories intended for ventilatory support equipment (intended only to augment the ventilation of spontaneously breathing patients), which are given in ISO 80601‑2-79 and ISO 80601-2-80.

¾     obstructive sleep apnoea therapy ME equipment, which are given in ISO 80601‐2‐70.

¾     user-powered resuscitators, which are given in ISO 10651‐4.

¾     gas-powered emergency resuscitators, which are given in ISO 10651‐5.

¾     continuous positive airway pressure (CPAP) ME equipment.

¾     high‐frequency jet ventilators (HFJVs), which are given in ISO 80601-2-87.

¾     high‐frequency oscillatory ventilators (HFOVs)[44], which are given in ISO 80601-2-87.

NOTE 6      An EMS ventilator can incorporate high-frequency jet or high-frequency oscillatory ventilation-modes.

¾     respiratory high-flow therapy equipment, which are given in ISO 80601-2-90.

NOTE 7      An EMS ventilator can incorporate high-flow therapy operational mode, but such a mode is only for spontaneously breathing patients.

¾     oxygen therapy constant flow ME equipment.

¾     cuirass or “iron‐lung” ventilators.

Replaced standards (1)
International standard Cancelled
Medical electrical equipment - Part 2-84: Particular requirements for the basic safety and essential performance of ventilators for the emergency medical services environment

<p>This document applies to the <i>basic safety</i> and <i>essential performance</i> of an <i>EMS</i> <i>ventilator</i> in combination with its <i>accessories</i>, hereafter also referred to as <i>ME equipment:</i> </p> <ul> <li> intended for <i>patients</i> who need differing levels of support from artificial ventilation including <i>ventilator-dependent</i> <i>patients</i>;</li> <li> intended to be operated by a <i>healthcare professional operator</i>;</li> <li> intended for use in the <i>EMS environment</i>; and</li> <li> intended for invasive or non-invasive ventilation.</li> </ul> <p>NOTE 1 An <i>EMS</i> <i>ventilator </i>can also be used for transport within a <i>professional healthcare facility</i>.</p> <p>* An <i>EMS ventilator</i> is not considered to utilize a <i>physiologic closed loop-control system</i> unless it uses a physiological <i>patient</i> variable to adjust the ventilation therapy settings.</p> <p>This document is also applicable to those <i>accessories</i> intended by their <i>manufacturer</i> to be connected to the <i>ventilator</i> <i>breathing system</i>, or to an <i>EMS</i> <i>ventilator</i>, where the characteristics of those <i>accessories</i> can affect the <i>basic safety</i> or <i>essential performance</i> of the <i>EMS</i> <i>ventilator</i>. </p> <p>NOTE 2 If a clause or subclause is specifically intended to be applicable to <i>ME equipment</i> only, or to <i>ME systems</i> only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to <i>ME equipment</i> and to <i>ME systems</i>, as relevant.</p> <p><i>Hazards</i> inherent in the intended physiological function of <i>ME equipment</i> or <i>ME systems</i> within the scope of this document are not covered by specific requirements in this document except in IEC 60601-1:2005, 7.2.13 and 8.4.1.</p> <p>NOTE 3 Additional information can be found in IEC 60601-1:2005+AMD1:2012, 4.2.</p> <p>This document does not specify the requirements for the following:</p> <ul> <li> <i>ventilators</i> or <i>accessories</i> intended for <i>ventilator-dependent</i> <i>patients</i> in critical care applications, which are given in ISO 80601-2-12. </li> <li> <i>ventilators</i> or <i>accessories</i> intended for<i> ventilator-dependent patients</i> in the <i>home healthcare environment</i>, which are given in ISO 80601-2-72<sup>[3]</sup>. </li> <li> <i>ventilators</i> or <i>accessories</i> intended for anaesthetic applications, which are given in ISO 80601‐2‐13<sup>[4]</sup>. </li> <li> <i>ventilators</i> or <i>accessories</i> intended for ventilatory support equipment (intended only to augment the ventilation of spontaneously breathing <i>patients</i>), which are given in ISO 80601‑2-79<sup>[5]</sup> and ISO 80601-2-80<sup>[6]</sup><sup> <sup>[1]</sup></sup>. </li> <li> obstructive sleep apnoea therapy <i>ME equipment</i>, which are given in ISO 80601‐2‐70<sup>[7]</sup>. </li> <li> <i>operator</i>-powered resuscitators, which are given in ISO 10651‐4<sup>[8]</sup>. </li> <li> gas-powered emergency resuscitators, which are given in ISO 10651‐5<sup>[9]</sup>. </li> <li> <i>continuous positive airway pressure</i> (<i>CPAP</i>) <i>ME equipment</i> .</li> <li> high‐frequency jet <i>ventilators</i> (HFJVs), which are given in ISO 80601-2-87<sup>[11]</sup>.</li> <li> high‐frequency oscillatory <i>ventilators</i> (HFOVs)<sup>[10]</sup>, which are given in ISO 80601-2-87<sup>[11]</sup>.</li> </ul> <p>NOTE 4 An <i>EMS</i> <i>ventilator</i> can incorporate high-frequency jet or high-frequency oscillatory <i>ventilation-modes</i>.</p> <ul> <li> cuirass or "iron‐lung" <i>ventilators</i>.</li> </ul> <p><sup><sup>[1]</sup></sup> ISO 80601-2-79 and ISO 80601-2-80 replace ISO 10651-6, which has been withdrawn.</p>

Need to identify, monitor and decipher standards?

COBAZ is the simple and effective solution to meet the normative needs related to your activity, in France and abroad.

Available by subscription, CObaz is THE modular solution to compose according to your needs today and tomorrow. Quickly discover CObaz!

Request your free, no-obligation live demo

I discover COBAZ