NF ISO 24236

NF ISO 24236

December 2006
Standard Current

Surface chemical analysis - Auger electron spectroscopy - Repeatability and constancy of intensity scale

ISO 24236:2005 specifies a method for evaluating the constancy and repeatability of the intensity scale of Auger electron spectrometers, for general analytical purposes, using an electron gun with a beam energy of 2 keV or greater. It is only applicable to instruments that incorporate an ion gun for sputter cleaning. It is not intended to be a calibration of the intensity/energy response function. That calibration may be made by the instrument manufacturer or other organization. The present procedure provides data to evaluate and confirm the accuracy with which the intensity/energy response function remains constant with instrument usage. Guidance is given on some of the instrumental settings that may affect this constancy.

View the extract
Main informations

Collections

National standards and national normative documents

Publication date

December 2006

Number of pages

22 p.

Reference

NF ISO 24236

ICS Codes

71.040.40   Chemical analysis

Classification index

X21-059

Print number

1 - 13/11/2006

International kinship

Sumary
Surface chemical analysis - Auger electron spectroscopy - Repeatability and constancy of intensity scale

ISO 24236:2005 specifies a method for evaluating the constancy and repeatability of the intensity scale of Auger electron spectrometers, for general analytical purposes, using an electron gun with a beam energy of 2 keV or greater. It is only applicable to instruments that incorporate an ion gun for sputter cleaning. It is not intended to be a calibration of the intensity/energy response function. That calibration may be made by the instrument manufacturer or other organization. The present procedure provides data to evaluate and confirm the accuracy with which the intensity/energy response function remains constant with instrument usage. Guidance is given on some of the instrumental settings that may affect this constancy.

Table of contents
View the extract
  • Avant-propos
    iv
  • Introduction
    v
  • 1 Domaine d'application
    1
  • 2 Symboles
    1
  • 3 Aperçu de la méthode
    2
  • 4 Méthode d'évaluation de la répétabilité et de la constance de l'échelle d'intensité
    4
  • 4.1 Obtention de l'échantillon de référence
    4
  • 4.2 Montage de l'échantillon
    4
  • 4.4 Choix des réglages du spectromètre pour lesquels la stabilité de l'intensité doit être déterminée
    4
  • 4.5 Fonctionnement de l'instrument
    5
  • 4.6 Options pour les mesures d'évaluation initiales ou ultérieures
    6
  • 4.7 Mesures de l'intensité et de la répétabilité
    6
  • 4.8 Calcul des intensités des pics, des rapports d'intensité et des incertitudes
    8
  • 4.9 Procédure d'évaluation régulière de la constance de l'échelle d'intensité
    9
  • 4.10 Évaluation suivante
    11
  • Annexe A (informative) Exemple de calculs et de mesures de la répétabilité de l'intensité pour un spectromètre d'électrons Auger commercial
    12
  • Bibliographie
    16