NF EN ISO 10993-6

NF EN ISO 10993-6

December 2009
Standard Cancelled

Biological evaluation of medical devices - Part 6 : tests for local effects after implantation

ISO 10993-6:2007 specifies test methods for the assessment of the local effects after implantation of biomaterials intended for use in medical devices.ISO 10993-6:2007 applies to materials that are:solid and non-biodegradable;degradable and/or resorbable;non-solid, such as porous materials, liquids, pastes and particulates.ISO 10993-6:2007 may also be applied to medical devices that are intended to be used topically in clinical indications where the surface or lining may have been breached, in order to evaluate local tissue responses.This part of ISO 10993 does not deal with systemic toxicity, carcinogenicity, teratogenicity or mutagenicity. However, the long-term implantation studies intended for evaluation of local biological effects may provide insight into some of these properties. Systemic toxicity studies conducted by implantation may satisfy the requirements of ISO 10993-6:2007.

View the extract
Main informations

Collections

National standards and national normative documents

Publication date

December 2009

Number of pages

32 p.

Reference

NF EN ISO 10993-6

ICS Codes

11.040.01   Medical equipment in general
11.100.20   Biological evaluation of medical devices

Classification index

S99-501-6

Print number

1 - 16/12/2009

International kinship

European kinship

EN ISO 10993-6:2009
Sumary
Biological evaluation of medical devices - Part 6 : tests for local effects after implantation

ISO 10993-6:2007 specifies test methods for the assessment of the local effects after implantation of biomaterials intended for use in medical devices.

ISO 10993-6:2007 applies to materials that are:

  1. solid and non-biodegradable;
  2. degradable and/or resorbable;
  3. non-solid, such as porous materials, liquids, pastes and particulates.

ISO 10993-6:2007 may also be applied to medical devices that are intended to be used topically in clinical indications where the surface or lining may have been breached, in order to evaluate local tissue responses.

This part of ISO 10993 does not deal with systemic toxicity, carcinogenicity, teratogenicity or mutagenicity. However, the long-term implantation studies intended for evaluation of local biological effects may provide insight into some of these properties. Systemic toxicity studies conducted by implantation may satisfy the requirements of ISO 10993-6:2007.

Replaced standards (1)
NF EN ISO 10993-6
August 2007
Standard Cancelled
Biological evaluation of medical devices - Part 6 : tests for local effects after implantation

Le présent document spécifie les méthodes d'essai pour l'évaluation des effets locaux après une implantation de biomatériaux destinés à être utilisés dans des dispositifs médicaux. Il s'applique aux matériaux solides et non biodégradables, dégradables et/ou résorbables et non solides ; comme les matériaux poreux, liquides, pâteux et particulaires. Il vient en appui des directives européennes 90/385/CEE sur les dispositifs médicaux implantables actifs et 93/42/CEE sur les dispositifs médicaux.

Standard replaced by (1)
NF EN ISO 10993-6
April 2017
Standard Current
Biological evaluation of medical devices - Part 6 : tests for local effects after implantation

ISO 10993-6:2016 specifies test methods for the assessment of the local effects after implantation of biomaterials intended for use in medical devices. ISO 10993-6:2016 applies to materials that are - solid and non-absorbable, - non-solid, such as porous materials, liquids, gels, pastes, and particulates, and - degradable and/or absorbable, which may be solid or non-solid. The test sample is implanted into a site and animal species appropriate for the evaluation of the biological safety of the material. These implantation tests are not intended to evaluate or determine the performance of the test sample in terms of mechanical or functional loading. This part of ISO 10993 can also be applied to medical devices that are intended to be used topically in clinical indications where the surface or lining might have been breached, in order to evaluate local tissue responses. The local effects are evaluated by a comparison of the tissue response caused by a test sample to that caused by control materials used in medical devices whose clinical acceptability and biocompatibility characteristics have been established. The objective of the test methods is to characterize the history and evolution of the tissue response after implantation of a medical device/biomaterial including final integration or absorption/degradation of the material. In particular for degradable/absorbable materials, the degradation characteristics of the material and the resulting tissue response should be determined. ISO 10993-6:2016 does not deal with systemic toxicity, carcinogenicity, teratogenicity or mutagenicity. However, the long-term implantation studies intended for evaluation of local biological effects might provide insight into some of these properties. Systemic toxicity studies conducted by implantation might satisfy the requirements of this part of ISO 10993. When conducting combined studies for evaluating local effects and systemic effects, the requirements of both standards is to be fulfilled.

Table of contents
  • Avant-propos
  • 1 Domaine d'application
  • 2 Références normatives
  • 3 Termes et définitions
  • 4 Dispositions communes concernant les méthodes d'essai d'implantation
  • 5 Méthodes d'essai, aspects généraux
  • 6 Rapport d'essai
  • Annexe C (normative) Méthode d'essai pour une implantation dans le tissu musculaire
  • Annexe D (normative) Méthode d'essai pour une implantation dans l'os
  • Annexe A (informative) Considérations générales relatives aux durées d'implantation et aux réponses tissulaires à des matériaux dégradables/résorbables
  • Annexe B (normative) Méthode d'essai pour une implantation dans le tissu sous-cutané
  • Annexe E (informative) Exemples d'évaluation des effets biologiques locaux après implantation
  • Annexe ZA (informative) Relation entre la présente Norme européenne et les exigences essentielles de la Directive européenne 93/42/CEE relative aux dispositifs médicaux
  • Annexe ZB (informative) Relation entre la présente Norme européenne et les exigences essentielles de la Directive européenne 90/385/CEE relative aux dispositifs médicaux implantables actifs
  • Bibliographie
ZOOM ON ... the Requirements department
To comply with a standard, you need to quickly understand its issues in order to determine its impact on your activity.

The Requirements department helps you quickly locate within the normative text:
- mandatory clauses to satisfy,
- non-essential but useful clauses to know, such as permissions and recommendations.

The identification of these types of clauses is based on the document “ISO / IEC Directives, Part 2 - Principles and rules of structure and drafting of ISO documents ”as well as on a constantly enriched list of verbal forms.

With Requirements, quickly access the main part of the normative text!

With Requirements, quickly access the main part of the normative text!
Need to identify, monitor and decipher standards?

COBAZ is the simple and effective solution to meet the normative needs related to your activity, in France and abroad.

Available by subscription, CObaz is THE modular solution to compose according to your needs today and tomorrow. Quickly discover CObaz!

Request your free, no-obligation live demo

I discover COBAZ