NF EN ISO 80601-2-12
Medical electrical equipment - Part 2-12 : particular requirements for basic safety and essential performance of critical care ventilators
This document applies to the basic safety and essential performance of a critical care ventilator in combination with its accessories, hereafter referred to as ME equipment: ¾ intended for use in an environment that provides specialized care for patients whose conditions can be life-threatening and who can require comprehensive care and constant monitoring in a professional healthcare facility; NOTE 2 For the purposes of this document, such an environment is referred to as a critical care environment. Ventilators for this environment are considered life-sustaining. NOTE 3 For the purposes of this document, such a critical care ventilator can provide ventilation during transport within a professional healthcare facility (i.e. be a transit-operable ventilator). NOTE 4 A critical care ventilator intended for use in transport within a professional healthcare facility is not considered as an emergency medical services environment ventilator.¾ intended to be operated by a healthcare professional operator; and ¾ intended for those patients who need differing levels of support from artificial ventilation including for ventilator-dependent patients. A critical care ventilator is not considered to use a physiologic closed-loop-control system unless it uses a physiological patient variable to adjust the artificial ventilation therapy settings.This document is also applicable to those accessories intended by their manufacturer to be connected to a ventilator breathing system, or to a ventilator, where the characteristics of those accessories can affect the basic safety or essential performance of the ventilator. NOTE 5 If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant.Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in IEC 60601-1:2005+AMD1:2012+AMD2:2020, 7.2.13 and 8.4.1.NOTE 6 Additional information can be found in IEC 60601-1:2005+AMD1:2012+AMD2:2020, 4.2.This document is not applicable to ME equipment or an ME system operating in a ventilator-operational mode solely intended for patients who are not dependent on artificial ventilation. NOTE 7 A critical care ventilator, when operating in such a ventilator-operational mode, is not considered life-sustaining.This document is not applicable to ME equipment that is intended solely to augment the ventilation of spontaneously breathing patients within a professional healthcare facility. This document does not specify the requirements for: NOTE 8 See ISO/TR 21954 for guidance on the selection of the appropriate ventilator for a given patient. ¾ ventilators or accessories intended for anaesthetic applications, which are given in ISO 80601‑2‑13;¾ ventilators or accessories intended for the emergency medical services environment, which are given in ISO 80601-2-84; ¾ ventilators or accessories intended for ventilator-dependent patients in the home healthcare environment, which are given in ISO 80601‑2-72; ¾ ventilators or accessories intended for home-care ventilatory support devices, which are given in ISO 80601-2-79 and ISO 80601-2-80; ¾ obstructive sleep apnoea therapy ME equipment, which are given in ISO 80601‑2‑70; ¾ continuous positive airway pressure (CPAP) ME equipment.¾ high-frequency ventilators, which are given in ISO 80601‑2‑87; NOTE 9 A critical care ventilator can incorporate high-frequency jet or high-frequency oscillatory ventilator-operational modes.¾ respiratory high-flow therapy equipment, which are given in ISO 80601‑2‑90; NOTE 10 A critical care ventilator can incorporate high-flow therapy operational mode, but such a mode is only for spontaneously breathing patients.¾ oxygen therapy constant flow ME equipment; and¾ cuirass or “iron-lung” ventilation equipment.
This document applies to the basic safety and essential performance of a critical care ventilator in combination with its accessories, hereafter referred to as ME equipment:
¾ intended for use in an environment that provides specialized care for patients whose conditions can be life-threatening and who can require comprehensive care and constant monitoring in a professional healthcare facility;
NOTE 2 For the purposes of this document, such an environment is referred to as a critical care environment. Ventilators for this environment are considered life-sustaining.
NOTE 3 For the purposes of this document, such a critical care ventilator can provide ventilation during transport within a professional healthcare facility (i.e. be a transit-operable ventilator).
NOTE 4 A critical care ventilator intended for use in transport within a professional healthcare facility is not considered as an emergency medical services environment ventilator.
¾ intended to be operated by a healthcare professional operator; and
¾ intended for those patients who need differing levels of support from artificial ventilation including for ventilator-dependent patients.
A critical care ventilator is not considered to use a physiologic closed-loop-control system unless it uses a physiological patient variable to adjust the artificial ventilation therapy settings.
This document is also applicable to those accessories intended by their manufacturer to be connected to a ventilator breathing system, or to a ventilator, where the characteristics of those accessories can affect the basic safety or essential performance of the ventilator.
NOTE 5 If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant.
Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in IEC 60601-1:2005+AMD1:2012+AMD2:2020, 7.2.13 and 8.4.1.
NOTE 6 Additional information can be found in IEC 60601-1:2005+AMD1:2012+AMD2:2020, 4.2.
This document is not applicable to ME equipment or an ME system operating in a ventilator-operational mode solely intended for patients who are not dependent on artificial ventilation.
NOTE 7 A critical care ventilator, when operating in such a ventilator-operational mode, is not considered life-sustaining.
This document is not applicable to ME equipment that is intended solely to augment the ventilation of spontaneously breathing patients within a professional healthcare facility.
This document does not specify the requirements for:
NOTE 8 See ISO/TR 21954 for guidance on the selection of the appropriate ventilator for a given patient.
¾ ventilators or accessories intended for anaesthetic applications, which are given in ISO 80601‑2‑13;
¾ ventilators or accessories intended for the emergency medical services environment, which are given in ISO 80601-2-84;
¾ ventilators or accessories intended for ventilator-dependent patients in the home healthcare environment, which are given in ISO 80601‑2-72;
¾ ventilators or accessories intended for home-care ventilatory support devices, which are given in ISO 80601-2-79 and ISO 80601-2-80;
¾ obstructive sleep apnoea therapy ME equipment, which are given in ISO 80601‑2‑70;
¾ continuous positive airway pressure (CPAP) ME equipment.
¾ high-frequency ventilators, which are given in ISO 80601‑2‑87;
NOTE 9 A critical care ventilator can incorporate high-frequency jet or high-frequency oscillatory ventilator-operational modes.
¾ respiratory high-flow therapy equipment, which are given in ISO 80601‑2‑90;
NOTE 10 A critical care ventilator can incorporate high-flow therapy operational mode, but such a mode is only for spontaneously breathing patients.
¾ oxygen therapy constant flow ME equipment; and
¾ cuirass or “iron-lung” ventilation equipment.
This document applies to the basic safety and essential performance of a ventilator in combination with its accessories, hereafter referred to as ME equipment: intended for use in an environment that provides specialized care for patients whose conditions can be life-threatening and who can require comprehensive care and constant monitoring in a professional healthcare facility; NOTE 1 For the purposes of this document, such an environment is referred to as a critical care environment. Ventilators for this environment are considered life-sustaining. NOTE 2 For the purposes of this document, such a ventilator can provide transport within a professional healthcare facility (i.e. be a transit-operable ventilator). NOTE 3 A critical care ventilator intended for use in transport within a professional healthcare facility is not considered as an emergency medical services environment ventilator. intended to be operated by a healthcare professional operator; and intended for those patients who need differing levels of support from artificial ventilation including for ventilator-dependent patients. A critical care ventilator is not considered to utilize a physiologic closed-loop-control system unless it uses a physiological patient variable to adjust the ventilation therapy settings. This document is also applicable to those accessories intended by their manufacturer to be connected to a ventilator breathing system, or to a ventilator, where the characteristics of those accessories can affect the basic safety or essential performance of the ventilator. NOTE 4 If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant. Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in IEC 60601-1:2005, 7.2.13 and 8.4.1. NOTE 5 Additional information can be found in IEC 60601-1:2005+AMD1:2012, 4.2. This document is not applicable to ME equipment or an ME system operating in a ventilator-operational mode solely intended for patients who are not dependent on artificial ventilation. NOTE 6 A critical care ventilator, when operating in such a ventilator-operational mode, is not considered life-sustaining. This document is not applicable to ME equipment that is intended solely to augment the ventilation of spontaneously breathing patients within a professional healthcare facility. This document does not specify the requirements for: ventilators or accessories intended for anaesthetic applications, which are given in ISO 80601-2-13[2]; ventilators or accessories intended for the emergency medical services environment, which are given in ISO 80601-2-84[3], the future replacement for ISO 10651-3[4]; ventilators or accessories intended for ventilator-dependent patients in the home healthcare environment, which are given in ISO 80601‑2-72:2015[5]; ventilators or accessories intended for home-care ventilatory support devices, which are given in ISO 80601-2-79:2018[6] and ISO 80601-2-80:2018[7][1]; obstructive sleep apnoea therapy ME equipment, which are given in ISO 80601‑2‑70[9]; continuous positive airway pressure (CPAP) ME equipment; high-frequency jet ventilators (HFJVs) and high-frequency oscillatory ventilators (HFOVs), which are given in ISO 80601‑2‑87[63]; NOTE 7 A critical care ventilator can incorporate high-frequency jet or high-frequency oscillatory ventilator-operational modes. oxygen therapy constant flow ME equipment; and cuirass or "iron-lung" ventilation equipment. [1] ISO 80601-2-79 and ISO 80601-2-80 replace ISO 10651-6, which has been withdrawn.
This document applies to the basic safety and essential performance of a ventilator in combination with its accessories, hereafter referred to as ME equipment: intended for use in an environment that provides specialized care for patients whose conditions can be life-threatening and who can require comprehensive care and constant monitoring in a professional healthcare facility; NOTE 1 For the purposes of this document, such an environment is referred to as a critical care environment. Ventilators for this environment are considered life-sustaining. NOTE 2 For the purposes of this document, such a ventilator can provide transport within a professional healthcare facility (i.e. be a transit-operable ventilator). NOTE 3 A critical care ventilator intended for use in transport within a professional healthcare facility is not considered as an emergency medical services environment ventilator. intended to be operated by a healthcare professional operator; and intended for those patients who need differing levels of support from artificial ventilation including for ventilator-dependent patients. A critical care ventilator is not considered to utilize a physiologic closed-loop-control system unless it uses a physiological patient variable to adjust the ventilation therapy settings. This document is also applicable to those accessories intended by their manufacturer to be connected to a ventilator breathing system, or to a ventilator, where the characteristics of those accessories can affect the basic safety or essential performance of the ventilator. NOTE 4 If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant. Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in IEC 60601-1:2005, 7.2.13 and 8.4.1. NOTE 5 Additional information can be found in IEC 60601-1:2005+AMD1:2012, 4.2. This document is not applicable to ME equipment or an ME system operating in a ventilator-operational mode solely intended for patients who are not dependent on artificial ventilation. NOTE 6 A critical care ventilator, when operating in such a ventilator-operational mode, is not considered life-sustaining. This document is not applicable to ME equipment that is intended solely to augment the ventilation of spontaneously breathing patients within a professional healthcare facility. This document does not specify the requirements for: ventilators or accessories intended for anaesthetic applications, which are given in ISO 80601-2-13[2]; ventilators or accessories intended for the emergency medical services environment, which are given in ISO 80601-2-84[3], the future replacement for ISO 10651-3[4]; ventilators or accessories intended for ventilator-dependent patients in the home healthcare environment, which are given in ISO 80601‑2-72:2015[5]; ventilators or accessories intended for home-care ventilatory support devices, which are given in ISO 80601-2-79:2018[6] and ISO 80601-2-80:2018[7][1]; obstructive sleep apnoea therapy ME equipment, which are given in ISO 80601‑2‑70[9]; continuous positive airway pressure (CPAP) ME equipment; high-frequency jet ventilators (HFJVs) and high-frequency oscillatory ventilators (HFOVs), which are given in ISO 80601‑2‑87[63]; NOTE 7 A critical care ventilator can incorporate high-frequency jet or high-frequency oscillatory ventilator-operational modes. oxygen therapy constant flow ME equipment; and cuirass or "iron-lung" ventilation equipment. [1] ISO 80601-2-79 and ISO 80601-2-80 replace ISO 10651-6, which has been withdrawn.
-
201.1 Domaine d'application, objet et normes collatérales
-
201.2 Références normatives
-
201.3 Termes et définitions
-
202 Perturbations électromagnétiques - Exigences et essais
-
206 Aptitude à l'utilisation
-
208 Exigences générales, essais et guide pour les systèmes d'alarme des appareils et des systèmes électromédicaux
- C Guide des exigences en matière demarquage et d'étiquetage des appareils EM et systèmes EM
- D Symboles des marquages
- AA Recommandations particulières et justifications
- BB Interfaces de données
- CC Référence auxprincipes essentiels et aux lignes directrices en matière d'étiquetage de l'IMDRF
- DD Référence auxprincipes essentiels
- Index alphabétique des termes définis
- Bibliographie
The Requirements department helps you quickly locate within the normative text:
- mandatory clauses to satisfy,
- non-essential but useful clauses to know, such as permissions and recommendations.
The identification of these types of clauses is based on the document “ISO / IEC Directives, Part 2 - Principles and rules of structure and drafting of ISO documents ”as well as on a constantly enriched list of verbal forms.
With Requirements, quickly access the main part of the normative text!

At a glance, you will be able to identify the additions, deletions or modifications to a text, table, figure and formula.

The Redlines + service is offered to you on the collection of French standards in force, in French language and in HTML and PDF format.
For an overview of the service, click on View a standard in redline format
COBAZ is the simple and effective solution to meet the normative needs related to your activity, in France and abroad.
Available by subscription, CObaz is THE modular solution to compose according to your needs today and tomorrow. Quickly discover CObaz!
Request your free, no-obligation live demo
I discover COBAZ